M70XP Moisture Monitor
The Model M70XP is designed to monitor moisture/water content in natural gas and biogas applications.
Range: 0–20lbs/mmscf
The Model M70XP is designed to monitor moisture/water content in natural gas and biogas applications. A heated conductive polymer sensor is utilized to provide a linear output of moisture concentration. The M70XP also offers an integrated sample conditioning system and Auto-Zero function that zeros the sensor on a daily or other configurable schedule.
Power | 12–24 VDC @ less than 2 watts |
Electrical classification | Certified for Class I, Div 1, Groups B, C & D |
Ambient | 0 to 60°C, 32 to 140°F (standard) Consult factory for other requirements |
Output Ranges | 0–20 lbs/mmscf standard 10–300 ppm (optional) (0.5 to 300 lbs/mmscf) |
Response Time | 20 seconds to 90% of step change wetting 300 seconds to 90% dry down |
Accuracy | 2% of reading or 10 ppm (0.5 lbs/million), whichever is greater |
Outputs | 4–20 mA Serial R2 232 and 485 Dual 3 amp solid state alarm relays Modbus |
Displays | Dual-line, 16-character display Moisture in ppm or lbs/million, sensor temperature °C, heater duty, calibration factor, and raw millivolts can be scrolled from internal button |
Auto Zero | Includes ASCO solenoid and extra flow meter. Inlet sample is dried daily (configurable) for 15 minutes in a Silica Gel Moisture trap at 80 psig to provide a daily repeatable moisture concentration reference, typically .5 lbs/mmscf. |
The Model M70XP is designed to monitor moisture/water content in natural gas and biogas applications.
The Model M90XP is designed to monitor percent level CO2 concentrations in natural gas and biogas applications.
The Model M80XP is designed to monitor CH4 in natural gas and biogas applications.
The Model M60XP is designed to monitor PPM and percent level O2 concentrations in natural gas and biogas applications.