Model M70XP

Moisture Monitor

Range: 0.5 – 20lbs/mmscf

The Model M70XP is designed to monitor moisture/water content in natural gas and biogas applications. A heated conductive polymer sensor is utilized to provide a linear output of moisture concentration. The M70XP also offers an integrated sample conditioning system and Auto-Zero function that zeros the sensor on a daily or other configurable schedule.

 

  • Features
    • Class I, Division 1
    • Range: 10 - 300 ppm; 0.5–20lbs/mmscf (custom ranges available)
    • Affordable measurement
    • Unit will display in lbs/mmscf and/or ppm
    • 2% of reading or 10 ppm (0.5 lb per million) accuracy
    • 4–20 mA output and alarms (2-wire)
    • RS-232 and RS-485 Serial Communications
    • Ultra-low power: Less than 2 watts power consumption
    • Dual-line, 16-character display
    • Daily Auto-Zero
  • Specifications
    Power

    12–24 VDC @ less than 2 watts (100 - 240 VAC Optional)

    Electrical classification Certified for Class I, Div 1, Groups B, C & D
    Ambient 0 to 50°C, 32 to 122°F (standard)
    Consult factory for other requirements
    Output Ranges 0.5 – 20 lbs/mmscf standard
    or 11 - 420 ppm
    (other options available upon request)
    Response Time 20 seconds to 90% of step change wetting
    300 seconds to 90% dry down
    Repeatability +/- 0.5 lbs/mmscf within range of 0.5 - 10 lbs/mmscf
    +/- 1lbs/mmscf within range of 10 - 20 lbs/mmscf
    Outputs 4–20 mA (concentration only)
    Serial R2 232 and 485 (all parameters)
    One (1) powered alarm relay (12V 3A)
    Modbus (optional)
    Displays Dual-line, 16-character display
    Line 1: Moisture in ppmv or lbs/mmscf
    Line 2: Alarm state, Heater duty, Offset, Raw mV, Alarm set point(s), Temperature °C (°F optional), manual zero
    Auto Zero Includes ASCO solenoid. Inlet sample is dried daily (configurable) for 15 minutes in a Moleseive Moisture trap at 10 - 15 psig to provide a zeroing of the sensor.
    Data Logging 6500 data points recorded every 10 minutes standard
    Note Product specifications subject to change without notice to improve reliability, function, design, or otherwise.
  • Downloads
  • Certification

Contact us to learn more about this product

M-Series Monitors

M70XP Moisture Monitor

The Model M70XP is designed to monitor moisture/water content in natural gas and biogas applications.

M90XP Carbon Dioxide Monitor

The Model M90XP is designed to monitor percent level CO2 concentrations in natural gas and biogas applications.

M80XP Methane Monitor

The Model M80XP is designed to monitor CH4 in natural gas and biogas applications.

M60XP Oxygen Monitor

The Model M60XP is designed to monitor PPM and percent level O2 concentrations in natural gas and biogas applications.