FREQUENTLY ASKED QUESTIONS

Ideal Cal-Gas concentration would at the value normally operating at or alarming at. Alternatively, Cal-Gas to use is 3/4 concentration of the range of the analyzer.
Calibration checks should be done every 3 to 6 months. Or as required by site preventive maintenance schedule.
The Filter should be changed as required. The frequency is dependent on the liquid content and particulates in the gas. A higher volume of liquids and particulates will require more frequent changes. Filters should be checked when analyzer is calibrated.
The standard H2S analyzer vents 4 standard cubic feet per hour.
The sweep valve should be just opened slightly until flow can be felt from the tubing. This will reduce lag time and allow liquids to be removed.
The analysis time is dynamic and changes with the concentration of the H2S during the sample.
Aperture
H2S (ppm)
Tape Life (Days)*
None
4.0 ppm
34.7
1/16th
16.0 ppm
30.4
*Tape life based on constant concentration sample over a averaged analysis time with a 300 ft tape.
By default the H2S analyzer does NOT time out when placed in bypass. A time out feature is available in the software, as well as Bypass Disable.
As long as there is some fluid in the humidifier it will be able to maintain the humidity of the gas. The fluid should be topped up each time the analyzer is calibrated or the tape is changed.
4-20mA wiring
The analog output of the H2S analyzer is a 4-20 mA signal calibrated at factory. The connections requires loop power from the PLC. If loop power is not able to be supplied, a powered AO board addition can be purchased.
Symptom
Solutions
PLC Indicating 0 mA
Wiring to analyzer is incorrect.
Power to loop is not present. If loop power not supplied then Powered AO board option may be required.
PLC Indicating ~2 mA
Analyzer in Bypass
PLC Indicating higher or lower signal than analyzer display
Scaling of analyzer is different sett
Determining the range
The range of the analyzer is set at the time of purchase. The accuracy and tape life are optimized to the customers request. The optimized range can be determined by the calibration sheet found in the back of the manual.

The Envent H2S analyzer is capable of indicating 10x the optimized range of H2S concentration. Accuracy of the reading can not be guaranteed when indicating outside the optimized range.
Optimizing the range
Range (ppm)
Aperture
0-10
None
0-20 *
1/16th
0-50
1/32nd
0-100
* If the shut in alarm value is less then 10 ppm no aperture is ideal.

For dilution analyzer systems the range adjustment depends on the dilution tubing. Please consult Envent about dilution range adjustments.
The sensor Fault is a result of the sensor mV level being below 500 mV or above 1200 mV. Most often caused by an incorrect zero being achieved.
  1. Put the Analyzer in Bypass.
  2. Turn off sample gas at flowmeter.
  3. Press the Menu/Set button until MTR RUN displayed.
  4. Press Left or Right on the Display to move the tape to a new white piece of tape.
  5. Remove the cover the sensor block. Press the small pushbutton located on the lower left side next to the wire connector. The sensor block will enter a “re-zero” procedure indicated by a red led.
  6. When the “re-zero”procedure is complete the green led should illuminate.
  7. Initiate another motor run.
  8. Press the Menu/Set button until zero is displayed. Value should be between 900 & 1100 mV.
  9. Turn on sample gas to flow of two.
  10. Take analyzer out of Bypass.
Gain numbers based on standard H 2S analyzers. Numbers do not apply to Total Sulfur and Dilution systems. Please consult factory cal-sheet for calibration numbers.
ProblemPossible ReasonsSolutions
Erratic H2S Readings
Trigger slide not seated properlyEnsure Trigger slid is seated in the groove of the sample chamber
Pressure in building moving up and down from fan, exhaust or windEductor may be plugged or bent blocked. Check that all vent tubing and fittings are 316 stainless steel, sized 3/8″ or larger.
Sample vent blocked or insufficient.Vent should be 1/2″ tubing on a downward slope. Ending at shepards hook. Possible heat trace required on tubing
Liquid carry over in sample conditioning system
Sample conditioning system requires cleaning. Refer to A-H 2S-Cleaning-Instructions_Nov-2011
Regulator not maintaining 15 psigPre-regulation to 50 psig of sample at sample point. Possible regulator requires repair or replacement.

Hydrocarbon liquid carried over through the sample regulator. Heated regulator may be required.

Contaminants in sample systemSample system has been flooded or contaminated. Sample system requires cleaning. Refer to Cleaning Procedure.
Analog Input 2 jumper removedRe-install jumper in Analog Input across (+4-20 & -4-20).
Sensor Block FaultRe-zero Sensor block. Refer to Sensor Re-Zero Procedure. Check for green light on sensor block.
Contaminants in sample chamberClean sample chamber. Replace aperture and window if required. Clear grease from window.

Envent no longer uses grease on chamber window. Applies to early H2S analyzers

Aperture Strip became dislodgedReattach aperture strip with RTV 102 silicone.
Tape Doesn’t AdvanceNo tension on take up reel.Check setscrew in take up reel collars. Check to see if manual advance is possible on tape
Tape BreakingHigh liquid content in sample gasGenie probe and additional filtration may be required
Feed wheel not spinning freelyDust and refuse build up between feed wheel and chassis. Requires removal and cleaning of chassis
Tape cover wheels pressing against tape.Tape cover wheel became warped. Needs to be flattened to not contact tape when on feed wheel bolt.
Trigger slide not seated properlyEnsure trigger slide is seated in groove of sample chamber.
Slow ResponseIncorrect Aperture for rangeRemoval or change of aperture type required. Contact Envent Tech support for ideal aperture size.
Liquid contamination in sample tubingSample conditioning system requires cleaning. Refer to Cleaning Procedure
Sensor Block FaultRe-Zero Sensor block. Refer to sensor re-zero procedure
Higher or Lower than expected readingLiquid contamination in sample tubingSample condition system requires cleaning. Refer to Cleaning Procedure.
Sample vent blockageVent should be 3/8″ or larger 316 stainless steel tubing on a downard slope. Heat trace maybe required to prevent freezing
Contaminants in sample chamberSample conditioning system requires cleaning. Cleaning Procedure
Fault light indicationSensor Low FaultRe-zero Sensor block. Refer to Sensor Re-zero Procedure
Sensor High faultRe-zero Sensor block. Refer to Sensor Re-zero Procedure
Low tapeSensing tape requires change. Possible low tape sensor failure
Low PressurePressure regulator outlet lower then setpoint of pressure switch. (factory set to 10 psi)
  • A-H2S-Cleaning-Instructions_Nov-2011
  • Venting
Have a question you don't see answered? Please contact us.