Environmental Temperature
- -20° to 60° C (-4° to 140° F) Quoted per application
Dimensions
- 32″ H x 14″ W x 8″ D (no sample system) (81.3 cm H x 35.6 cm W x 20.3 cm D)
Electrical Classification
- Class I, Division 2, Groups B, C, and D
Power
- 120 VAC 50/60 Hz Standard
- 240 VAC 50/60 Hz Available
Power Consumption
- Start up: 100 watts (does not include sample system electronics)
- Steady State: 60 - 80 watts nominal
GC Valves
- Six-port and ten-port diaphragm chromatograph valves (application dependent)
Carrier Gas
- UHP Helium (99.999%) or UHP Hydrogen (99.999%)
Actuation Gas
- Helium, Hydrogen, Instrument Air (GC Valves/Stream Valves Regulated to 65 psig)
Detector
- Thermal conductivity detector (TCD): Single or Dual TCD capabilities
Input/Output
- Two (2) analog outputs
- Four (4) dry contact relay outputs
- Four (4) digital inputs
- Four (4) solenoid outputs
Communications
- SIM 2251 Modbus Mapping
- User Modbus Mapping
- One (1) RS-232 serial communication port (Modbus capable)
- Two (2) RS-485 serial communication ports (Modbus capable)
- One (1) ethernet communication port RJ-45 (Modbus capable)
Measurement Calculations
- Latest GPA 2145, GPA 2172, AGA 8, and ISO 6976 calculations