Model 132S BTU Gas Chromatograph
Envent's Class I Division 2 GC is designed to provide accurate and reliable measurements for a variety of applications.
For Gas Processing Applications
The Model 132S Process Gas Chromatograph (GC) is a simple approach to energy measurement, created and designed for many different applications.
Envent provides a Process Gas Chromatograph platform that is efficiently manufactured to ensure industry leading delivery, while providing a GC that allows for ease of serviceability.
Typical applications include:
Ambient Temperature | -20°C to 60°C (-4°F to 140°F). |
Dimensions | 32″ H x 14″ W x 8″ D (no sample system), (81.3cm H x 35.6cm W x 20.3cm D). |
Mounting | Wall mount or Floor mount |
Weight | Approximately 40kg (88 lbs.). |
Electrical classification | Certified for Class I, Division 2, Groups B, C, D |
Power | 120 VAC 50/60 Hz Standard, 240 VAC 50/60 Hz Available |
Power Consumption | Start up: 175 watts. Steady State: 100 watts nominal |
Oven | Airless Heat Sink Iso-Thermal, maximum 150°C (302°F). |
Valves | Six-port and ten-port diaphragm chromatograph valves. Liquid injection rotary valves |
Performance/Repeatability |
C6+ 4 minute Controlled Temperature: ±0.25 Btu per 1,000 Btu/SCF (±0.025%) at ambient; ±0.4 Btu per 1,000 Btu (±0.04%) over temperature range; of -5°F to 140°F (-20°C to 60°C) 20°C to 60°C (-4° to 140°F) +/- 0.4 Btu per 1,000 Btu/SCF C6+ 2 minute Controlled Temperature: +/- 0.5 Btu, per 1,000 Btu/SCF -20° to 60°C (-4° to 140°F) +/- 1 Btu per 1,000 Btu/SCF |
Carrier Gas |
Application-dependent. Typically UHP Helium (99.995%), Hydrogen or Nitrogen available upon request |
Detector | Thermal conductivity detector (TCD). |
Gating Options | Fixed-time, auto-detection. |
Streams | Up to 10 streams (includes calibration streams). |
Communications | *Two analog outputs. *Four dry contact relay outputs. Four digital inputs. Eight digital outputs. Two RS-232 serial communication ports. Four RS-485 serial communication ports. One Ethernet communications port. *Eight additional analog outputs and four additional dry contact relays with Analog Expander Board (optional). |
GC Software | Windows® software for GC configuration, archive retrieval, and Modbus mapping. |
Custody Transfer | Approved for Custody Transfer Applications by Measurement Canada. |
Note | Product specifications subject to change without notice to improve reliability, function, design, or otherwise. |
Envent's Class I Division 2 GC is designed to provide accurate and reliable measurements for a variety of applications.
Envent's Class I Division 1 GC is designed to provide accurate and reliable measurements for a variety of applications.
Envent's Class I Division 2 GC is designed to provide accurate and reliable measurements for a variety of applications.
Envent's Class I Division 1 GC is designed to provide accurate and reliable measurements for a variety of applications.